Z-2122-2025: VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides

Recall record Z-2122-2025 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2122-2025
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jul 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Source facts

Action
On June 10, 2025, "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letters were sent to distributors and customers. CUSTOMER LETTERS: REQUIRED ACTIONS " Monitor VITROS CHE Slides QC results to detect a drift in VITROS Performance Verifiers I and II below 2x Within-lab SD limits. If QC results are below 2x WLSD, perform QC again using freshly reconstituted vials of VITROS Performance Verifiers I and II. " Complete the enclosed Confirmation of Receipt form no later than June 18, 2025. " Save this notification with your User Documentation or post this notification near your laboratory's fluid reconstitution location until the issue has been resolved. " Please forward this notification if the affected products were distributed outside of your facility. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization. QuidelOrtho's investigation is ongoing, and we are working to identify root cause. We will communicate again after root cause has been determined and the issue has been resolved. Please note that until this issue is resolved, future lots of VITROS Performance Verifiers I and II may not meet the 7-day reconstituted stability for CHE. DISTRIBUTORS: REQUIRED ACTIONS " Please send the enclosed customer letter and customer Confirmation of Receipt form to all customers who have been shipped VITROS Chemistry Products Performance Verifiers AND VITROS CHE Slides from August 1, 2023, to date. " Complete the enclosed distributor Confirmation of Receipt form no later than June 18, 2025. " Please forward this notification if the affected products were distributed outside of your facility. If you have further questions, please contact the Global Services Organization at 1-800-421-3311.
Initiated
2025-06-10
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
Recall number
Z-2122-2025
Recall status
Open, Classified
CDRH record ID
214350
FDA classified
2025-07-17
Recall event number
97047

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗