Open original FDA record ↗Recall record
- FDA identifier
- Z-2123-2017
- Company in FDA record
- Merge Healthcare, Inc.
- FDA classification date
- May 22, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
Source facts
- Action
- The recalling firm issued a "Medical Device Correction" letter dated 4/13/2016 via e-mail. The letter identified the affected product, issue involved, potential harm, a work around, as well as the actions to be performed by both the customer and firm. Merge Technical Support will contact customers to discuss the fix. Customers are to ensure that all users are provided the notice. If the affected product was further distributed, those customers should be identified and notified of the issue. A copy of the notice may be provided to those customers. If there are additional questions, customers can e-mail recall@merge.com.
- Initiated
- 2016-04-13
- Root cause
- Software design
- Terminated
- 2017-07-17
- Distribution
- Distributed to the states of AZ, CT, FL, GA, IL, IN, LA, MI, NY, NC, OH, OK, TX, and VT.
- Recall number
- Z-2123-2017
- Recall status
- Terminated
- CDRH record ID
- 154998
- FDA classified
- 2017-05-22
- Recall event number
- 77124