Skip to content

Public FDA record

Z-2125-2017: Legacy Full Contour Laboratory Abutment: 3.5mmD Platform/5.5mmD

Recall record Z-2125-2017 for Implant Direct Sybron Manufacturing, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2125-2017
Company in FDA record
Implant Direct Sybron Manufacturing, LLC
FDA record creation date
May 22, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Implant Direct Sybron Manufacturing is recalling the Legacy Full-Contour Abutment 3.5mmD Assembly because it may be out of Implant Direct specification.

Source facts

Action
Implant Direct sent an Urgent Medical Device Field Corrective recall letter dated April 2017 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are informed that the product may be out of Implant Direct specification. The abutment should have a full hex configuration from end of part to shoulder (1.6mm). When the part is in-correct there will be a noticeable diameter ring just above shoulder and a hex flat face that appears short (@ 1.1mm) not allowing full engagement to implant analog. This discrepancy may lead to the possibility of the abutment not seating properly on the implant, and if undetected may lead to micro-leakage and a possible infection to the patient. Customers are instructed to complete and return the Acknowledgement and Recall Return form within 48 hrs. Customers with any questions are instructed to contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425.
Initiated
2017-05-02
Root cause
Nonconforming Material/Component
Terminated
2018-01-08
Distribution
Worldwide Distribution - Nationwide Distribution and to the countries of : European Union (HU, GB, DE, DK, IT, HR, FR), and Japan
Recall number
Z-2125-2017
Recall status
Terminated
CDRH record ID
155149
FDA record created
2017-05-22
Recall event number
77171

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗