Skip to content

Public FDA record

Z-2126-2011: Cirrus HD-OCT Model 400 instruments; new and refurbished, with or upgraded to, version 4.6 or 5.1.0 softw

Recall record Z-2126-2011 for Carl Zeiss Meditec Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2126-2011
Company in FDA record
Carl Zeiss Meditec Inc
FDA classification date
May 2, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A software defect exists in versions 5.0.0 and 5.1.0 where when following a specific workflow, exam data may be saved under the name of the first patient

Source facts

Action
Carl Zeiss Meditec, Inc. sent an Field Correction Action letter dated March 9, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letters requested acknowledgement of the letter and a response and the fact that a follow up packet with revised software will be sent to all customers with V 5.0.0 or 5.1.0 software. For questions, please call 1-855-853-9461.
Initiated
2011-03-10
Root cause
Software design
Terminated
2012-03-29
Distribution
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, and WV, and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Boliva, Brazil, Canada, France, Chili, China, Columbia, Denmark, Ecuador,Germany, Great Britain, Greece, Guatemala, Hong Kong, India, Japan, Lebanon, Malaysia, Mexico, New Zealand, Pakistan, Peru, Singapore,South Korea, Spain, Sri Lanka, Taiwan, Uruguay and Venezuela
Recall number
Z-2126-2011
Recall status
Terminated
CDRH record ID
98969
FDA classified
2011-05-02
Recall event number
58313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.