Skip to content

Public FDA record

Z-2126-2025: Brand Name: EMPOWR 3D KNEE TIBIAL INSERT - E-PLUS Product Name: EMPOWR 3D KNEETM INS, 11L 12MM, VE Model/

Recall record Z-2126-2025 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2126-2025
Company in FDA record
Encore Medical, LP
FDA classification date
Jul 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. The use of this product may result in the incorrect product being selected. This could result in a revision surgery if used.

Source facts

Action
On 06/02/2025, the firm emailed an "Urgent: Medical Device Recall" letter to customers. The letter described the product, problem and actions to be taken. Customers are instructed to: 1.Review their stock for the products and lot numbers for the items to be returned. 2.This notice needs to be passed on to all those within their organization or to any organization where the product has been transferred. 3.Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through their regional email address, to place a replacement request. They will be provided with an RMA number and shipping label to return the affected product. 4.Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. Any questions, contact Recall Coordinator at productsafety@enovis.com or call/text 682-299-9133.
Initiated
2025-06-02
Root cause
Packaging change control
Distribution
U.S Distribution to states of: AZ, IA, ID, IL, IN, LA, MI, MO, MS, NE, NY, OK, OR, PR, SD, UT, VA, and WA
Recall number
Z-2126-2025
Recall status
Open, Classified
CDRH record ID
214353
FDA classified
2025-07-17
Recall event number
97048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗