Z-2127-2025: Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following

Recall record Z-2127-2025 for Medtronic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2127-2025
Company in FDA record
Medtronic, Inc.
FDA classification date
Jul 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.

Source facts

Action
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTIFICATION to its consignees on 06/18/2025 via UPS 2nd day delivery. The notice announced that a software update was available for the Medtronic CareLink SmartSync Device Manager (SmartSync), explained the issue which the update was correcting, risk, and requested the following: Customer Actions: " Update SmartSync (to version 4.2.3 or higher). Refer to release notes provided by your local Medtronic representative for update instructions. If needed, Medtronic representatives are available to work with you to install or update the SmartSync application(s) on your tablet. See Appendix for instructions on how to verify the software update is complete. " Sign and return the enclosed confirmation form and share this notice with those who need to be aware within your organization or with any organization where SmartSync may be in use. Questions regarding this information should be directed to the local Medtronic Representative or Medtronic Programmer Technical Services at 800-638-1991, Option 1.
Initiated
2025-06-17
Root cause
Software design
Distribution
Worldwide
Recall number
Z-2127-2025
Recall status
Open, Classified
CDRH record ID
214616
FDA classified
2025-07-17
Recall event number
97125

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗