Z-2128-2023: Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457

Recall record Z-2128-2023 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2128-2023
Company in FDA record
Abiomed, Inc.
FDA classification date
Jul 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally introduced during manufacturing at the joint between the sidearm and the hub may cause the joint to detach.

Source facts

Action
An URGENT: VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) notification letter was delivered to customers on 5/23/23. Customers were also contacted via telephone on 5/26/23. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1. Examine your inventory immediately to determine if you have product subject to this recall (removal). DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your Abiomed local clinical field staff to coordinate the return of the subject products. 3. Review, complete all fields, sign, and return the attached Business Response Form (BRF) on the last page of this letter to the Recall Coordinator identified in this document. 4. Forward this notice to any personnel in your facility who needs to be informed. 5. If any of the product subject to this recall (removal) has been forwarded to another facility, contact that facility immediately and provide a copy of this notice to the relevant personnel. 6. Post a copy of this field safety notice in a visible area for awareness. 7. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA s MedWatch Adverse Event Reporting Program using the link below: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reportingprogram/reporting-serious-problems-fda If you have questions or concerns regarding this notice, please contact (Shashi Thoutam) directly at +1(734) 262-6255 and/or your local clinical field staff.
Initiated
2023-05-26
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of MN, NJ, TN.
Recall number
Z-2128-2023
Recall status
Open, Classified
CDRH record ID
201284
FDA classified
2023-07-11
Recall event number
92521

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗