Open original FDA record ↗Recall record
- FDA identifier
- Z-2129-2011
- Company in FDA record
- Baxter Healthcare Renal Div
- FDA classification date
- May 3, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Clinicians are being reminded of the current warning in The HomeChoice APD Systems Patient At-Home Guide, 07-19-61-244 (October 2,2009), which states on page 3-13 "To reduce the risk of bums, electrocution, fire, or injury to persons: Close supervision is necessary when this product is used by, on, or near children or those unable to care for themselves." Tubing entanglement around the neck may r
Source facts
- Action
- Baxter sent Safety Alert letters dated March 15, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to consider the following safeguards for at-risk patients on HomeChoice Therapy: Have a caregiver present while the patient is performing therapy. Move the HomeChoice cycler to the foot of the bed, if possible, to minimize the likelihood of entanglement that may lead to asphyxiation. Install bed railings for patients at risk for seizures or falls from bed. Complete the attached customer reply form and fax it to Baxter at t847-270-5457. For questions regarding this recall call 888-736-2543, option 2.
- Initiated
- 2011-03-16
- Root cause
- Other
- Terminated
- 2012-03-08
- Distribution
- Worldwide Distribution - USA including AZ, CA, CT, DC, FL, KY, MO, NC, NM, Ny, OH, PA, TN, TX, and VA and the countries of Austria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Ireland, Israel, Italy, South Korea, Kuwait, Lebanon, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russia, South Africa, Saudi Arabia, Singapore, Slovania, Spain, Sweden, Switzerland, Thailand, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingom, and Venezuela.
- Recall number
- Z-2129-2011
- Recall status
- Terminated
- CDRH record ID
- 99244
- FDA classified
- 2011-05-03
- Recall event number
- 58383