Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
Source facts
Action
Medtronic sent a Field Safety Notice letter dated August 1, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Medtronic will perform an inspection on site to verify and correct the grounding connection of the PoleStar System. For questions contact Medtronic Navigation Technical Services at 1-800-595-9709.
Initiated
2016-06-20
Root cause
Process change control
Terminated
2017-06-21
Distribution
Worldwide Distribution - US Nationwide.
Recall number
Z-2129-2016
Recall status
Terminated
CDRH record ID
147636
FDA classified
2016-07-06
Recall event number
74464