Z-2129-2023: VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calib

Recall record Z-2129-2023 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2129-2023
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Jul 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.

Source facts

Action
QuidelOrtho (formerly Ortho Clinical Diagnostics) issued Urgent product correction Notification on 5/8/23. Letter states reason for recall, health risk and action to take: " Immediately discontinue using and discard your remaining inventory of VITROS Calibrator Kit 20, Lot 2022. QuidelOrtho will replace or credit your account. Indicate quantities to be replaced or credited on the attached Confirmation of Receipt form. " If you do not have an alternative lot of VITROS Calibrator Kit 20, you must discontinue reporting patient results using a Lot 2022 calibration. " Upon receipt of your replacement lot of VITROS Calibrator Kit 20, re-calibrate VITROS Transferrin, C3, C4, IgA, IgG and/or IgM assays on your VITROS System. " Complete the enclosed Confirmation of Receipt form . " Please forward this notification if the affected product was distributed outside of your facility. " Post this notification by your VITROS System(s) or with the user documentation until you receive your replacement order. Resolution Once we receive your completed Confirmation of Receipt form, QuidelOrtho will process your replacement order or credit your account. To provide replacement product for all customers, product allocation may be necessary. Contact Information . If you have further questions, please contact our Ortho Care Technical Solutions Center
Initiated
2023-05-08
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden , The Netherlands, United Kingdom.
Recall number
Z-2129-2023
Recall status
Open, Classified
CDRH record ID
200908
FDA classified
2023-07-11
Recall event number
92440

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗