Skip to content

Public FDA record

Z-2129-2025: Plum Duo Infusion System, List Number: 400020401

Recall record Z-2129-2025 for ICU Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2129-2025
Company in FDA record
ICU Medical, Inc.
FDA classification date
Aug 6, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.

Source facts

Action
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 7/11/25 was sent to customers. The letter described the product, problem and actions to be taken. Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. 2. To stop an infusion, ensure that users carefully attend to the "CONFIRM STOP" message that appears after pressing the STOP button. The "CONFIRM STOP" message includes a yes/no selection to confirm the intent to stop the infusion, thereby helping to prevent unintended interruptions. 3. If the pump becomes unresponsive, exchange the pump and reprogram. Follow-up Actions by ICU Medical: ICU Medical will develop a patch to correct this issue for Plum Duo. ICU Medical will contact you to schedule the software update for your pumps if needed. https://icumed.custhelp.com/app/market-action For any questions regarding this action Technical Support 1-(800)-241-4002, option 3 tsc.support@icumed.com
Initiated
2025-07-11
Root cause
Under Investigation by firm
Distribution
US Nationwide.
Recall number
Z-2129-2025
Recall status
Open, Classified
CDRH record ID
214769
FDA classified
2025-08-06
Recall event number
97186

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗