Open original FDA record ↗Recall record
- FDA identifier
- Z-2129-2025
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- Aug 6, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
Source facts
- Action
- An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 7/11/25 was sent to customers. The letter described the product, problem and actions to be taken. Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. 2. To stop an infusion, ensure that users carefully attend to the "CONFIRM STOP" message that appears after pressing the STOP button. The "CONFIRM STOP" message includes a yes/no selection to confirm the intent to stop the infusion, thereby helping to prevent unintended interruptions. 3. If the pump becomes unresponsive, exchange the pump and reprogram. Follow-up Actions by ICU Medical: ICU Medical will develop a patch to correct this issue for Plum Duo. ICU Medical will contact you to schedule the software update for your pumps if needed. https://icumed.custhelp.com/app/market-action For any questions regarding this action Technical Support 1-(800)-241-4002, option 3 tsc.support@icumed.com
- Initiated
- 2025-07-11
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide.
- Recall number
- Z-2129-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214769
- FDA classified
- 2025-08-06
- Recall event number
- 97186