Z-2130-2023: Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a whee

Recall record Z-2130-2023 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2130-2023
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA classification date
Jul 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.

Source facts

Action
Stryker issued Urgent Medical Device Correction Letter on May 24, 2023 via Fed'X. Letter states reason for recall, health risk and action to take: 1. Locate the units listed on the attached business reply form and identify the address where they can be serviced. 2. Review and follow the required customer actions set forth in Appendix A to ensure the unit continues to operate safely until the correction can be performed. If any of the issues listed are present, remove the unit from service and contact Tech Support to request a service visit. 3. Return the enclosed business reply form to confirm receipt of this notification by fax +1 269 488 8691 or email productfieldaction@stryker.com. 4. Stryker will be initiating a correction for all affected Power-PRO 2 units and will contact customers to schedule device corrections once the correction is ready for implementation. The company will provide customers with updated information when it is available. 5. If you have loaned or sold any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. 6. If you have disposed of any of these units and they are no longer in use, please, advise us of their obsolescence by providing us with their serial number in the space provided on the business reply form. If you have any questions or concerns, contact Technical Support at 1 800 787 9537 option 2 Monday through Friday from 8:00 a.m. to 6:00 p.m. ET.
Initiated
2023-05-24
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom.
Recall number
Z-2130-2023
Recall status
Open, Classified
CDRH record ID
201164
FDA classified
2023-07-12
Recall event number
92499

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗