Skip to content

Public FDA record

Z-2130-2024: ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Refer

Recall record Z-2130-2024 for ZOLL Medical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2130-2024
Company in FDA record
ZOLL Medical Corporation
FDA classification date
Jun 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy

Source facts

Action
Zoll issued Urgent Corrective Action Letter dated 4/22/24. Letter states reason for recall, health, risk and action to take: Customers who have affected manuals should take the following steps: (1) Locate all old versions of affected Operator s Guides and/or Quick Reference Guides. (2) Complete and return the attached customer response form via e-mail, fax, or mail to request replacement manuals. (3) Alert all users of the ZOLL Ventilator - with MRI Compatibility, to ensure they are aware of the safe distance requirements and are trained to these manual updates. (4) Dispose of obsolete guides. Technical support number are 1-800-348-9011 or 1-978-421-9460 and are available to assist users with any aspect of this notice
Initiated
2024-04-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Argentina, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Bolivia, Brazil, Canada, Switzerland, China, Colombia, Czech Republic, Germany, Spain, Great Britain (UK), Greece, Guatemala, Guam, Hong Kong, Croatia, Hungary, Indonesia, Israel, India, Iraq, Italy, Japan, South Korea, Lebanon, Myanmar, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Thailand, Turkey, Taiwan, Vietnam.
Recall number
Z-2130-2024
Recall status
Open, Classified
CDRH record ID
207637
FDA classified
2024-06-20
Recall event number
94589

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.