Z-2131-2016: GE Healthcare, Discovery MR750w

Recall record Z-2131-2016 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2131-2016
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Jul 6, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. No injury has been reported to GE as a result of this issue.

Source facts

Action
Consignees were sent on 6/8/2016 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 60901 dated June 8, 2016. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, and Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2016-06-08
Root cause
Nonconforming Material/Component
Terminated
2018-10-04
Distribution
US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY. OUS: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bermuda Brazil Bulgaria Canada China Egypt France Germany Hungary India Indonesia Ireland Italy Japan Kazakhstan Korea (Republic of) Kuwait Lebanon Mexico Netherlands New Zealand Poland Qatar Romania Russia Russian Federation Saudi Arabia Singapore Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom Uruguay
Recall number
Z-2131-2016
Recall status
Terminated
CDRH record ID
147567
FDA record created
2016-07-06
Recall event number
74442

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗