Z-2131-2020: da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

Recall record Z-2131-2020 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2131-2020
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Jun 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.

Source facts

Action
Urgent Medical Device Recall Letters (Field Safety Notices) were hand-delivered to U.S. customers on August 19, 2019 and OUS customers on August 20, 2019. Customers were asked to locate affected camera(s) in their inventory and inspect per the instructions attached to the Field Safety Notice throughout the life of the camera(s). If residual soil is present, customers were to return the affected camera(s) to the manufacturer using the standard Return Materials Authorization (RMA) process. Additionally, customers were requested to: iInform necessary hospital personnel about this issue; complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form; and to retain a copy of this letter and the Acknowledgement Form for their files. Customers were informed that an Intuitive Representative would schedule a site visit to review the attached inspection instructions, and that the firm will provide a replacement camera, at no charge, for returned affected product. Customers requiring further information or support may contact their Clinical Sales Representative or contact Intuitive Customer Service at the following numbers: US Customer Service (800) 876-1310, Option 3 (7 AM to 5 PM EST) or Customerservice@intusurg.com, or South Korea Customer Service: + 82-02-3271-3200, Option 3 ( 9AM to 6 PM KSTJ)
Initiated
2019-08-19
Root cause
Process control
Distribution
Distribution US nationwide and South Korea.
Recall number
Z-2131-2020
Recall status
Open, Classified
CDRH record ID
180523
FDA record created
2020-06-01
Recall event number
85324

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗