i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.
Source facts
Action
Abbott Point of Care, Inc sent a Urgent Recall Notice letter dated January 14, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were instructed to discontinue use of Cartridges from the identified lot. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card.
If you have any questions regarding this information please contact your Abbott Point of Care Technical Support at 800-366-8020 Option 1, or your Abbott Point of Care representative.
Initiated
2011-01-01
Root cause
Nonconforming Material/Component
Terminated
2013-05-22
Distribution
USA ( nationwide )
Recall number
Z-2133-2012
Recall status
Terminated
CDRH record ID
110354
FDA classified
2012-08-02
Recall event number
62371