There is a misassembly issue with the
photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by
Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Source facts
Action
An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
Initiated
2014-05-27
Root cause
Employee error
Terminated
2014-12-15
Distribution
FL
LA
AR
IA
OH
OH
WI
IL
LA
KS
MT
Recall number
Z-2133-2014
Recall status
Terminated
CDRH record ID
127514
FDA classified
2014-07-30
Recall event number
68400