Z-2134-2020: Philips Sterilizable Defibrillator Internal Paddles - Internal paddles are an accessory to the compatible

Recall record Z-2134-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2134-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
May 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Internal Paddles may wear over time and may not be safe or ready for use, unless the directions in the Instructions for Use (IFU) to perform routine operational checks are followed. The instructions for use rely on these checks rather than specifying a maximum number of sterilization cycles to determine when the paddles should be discarded. A damaged Internal Paddle may not be able to deliver therapy.

Source facts

Action
Philips issued U.S. customer Field Safety Notification on 04/20/2020 via priority mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S.. Letter states reason for recall, health risk an action to take: Follow the Instructions for Use, Paddles Checks section. The Paddle Checks activities include; Mechanical Check, Visual Inspection, Functional Check, and Continuity Check. Perform these activities to confirm the paddles are safe and ready for use. Continue to perform the Paddles Checks activities as recommended in the IFU before use as this reduces the risk of a failure. If one or more of these Paddles Checks activities fail, you must remove them from service and replace the paddles. To acknowledge receipt of this notification, please complete and fax the Customer Reply Form to: +1.877.499.7223 or email to recall .response@philips.com. If you need further information or support concerning this notification, please AND SUPPORT contact your local Philips representative or call us at 1-800-722-9377
Initiated
2020-04-20
Root cause
Device Design
Terminated
2021-04-02
Distribution
Worldwide distribution - US Nationwide distrubtion and the countries of Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Congo (Democratic Republic of the), Costa Rica, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Haiti, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Lao People's , Democratic Republic Latvia, Lebanon, Libya, Macao, Macedonia, Malaysia, Maldives, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman Pakistan, Palestine, State of Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania Russian Federation, Saudi Arabia, Serbia, Singapore Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.
Recall number
Z-2134-2020
Recall status
Terminated
CDRH record ID
181256
FDA record created
2020-05-27
Recall event number
85571

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗