Skip to content

Public FDA record

Z-2134-2021: CytoCell DiGeorge/VCFS TUPLE1 Region and 22q13.3 Region Probe, Model Number LPU004-A. analyte specific r

Recall record Z-2134-2021 for Cytocell Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2134-2021
Company in FDA record
Cytocell Ltd.
FDA record creation date
Jul 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device may show unexpected locus specific signals in addition to those at 22q. Users may observe faint additional locus specific signals at 20p12/13.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 5/17/21 was sent to customers. Actions to be taken by [Customer/User] Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signal at 20p12/13. Corrective and Preventative Actions: Cytocell are currently investigating how this issue occurred and will be implementing the necessary corrective and preventative actions to reduce the risk of recurrence. Your assistance is appreciated and necessary to ensure the interpretation of a result from a resulting LDT is not affected. Please complete and return the enclosed response form as soon as possible. We wish to sincerely apologise for any inconvenience caused as a result of this medical device recall. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Initiated
2021-05-17
Root cause
Nonconforming Material/Component
Terminated
2022-08-15
Distribution
Distribution to US states of CA, IL, MO, NY, PA, TX, and UT, and Canada.
Recall number
Z-2134-2021
Recall status
Terminated
CDRH record ID
187967
FDA record created
2021-07-23
Recall event number
88138

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗