Open original FDA record ↗Recall record
- FDA identifier
- Z-2135-2019
- Company in FDA record
- Siemens Healthcare Diagnostics Inc
- FDA record creation date
- Jul 30, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results.
Source facts
- Action
- On May 29, 2019, Siemens issued Urgent Medical Device Correction notices to U.S.A. customers via FedEx courier service and Urgent Field Safety Notices to international customers via email. Actions to be Taken by the Customer: 1. Do not use TestPaks within 2 mos of expiration. 2. Repeat samples with cTNL results above 0.07 ng/mL or your institution's established 99th percentile value. 3. Complete and return the Field Correction Effectiveness Check form 4. Retain a copy of the letter with your laboratory records and forward a copy of the notice to anyone to whom you may have distributed the product. Siemens is working to resolve this issue and will contact customers when additional information is available. Siemens will reimburse users for repeat tests and discarded cTNL TestPaks within 2 months of the printed expiration date associated with this field corrective action.
- Initiated
- 2019-05-29
- Root cause
- Under Investigation by firm
- Terminated
- 2022-05-12
- Distribution
- US Nationwide Distribution
- Recall number
- Z-2135-2019
- Recall status
- Terminated
- CDRH record ID
- 173293
- FDA record created
- 2019-07-30
- Recall event number
- 82990