Open original FDA record ↗Recall record
- FDA identifier
- Z-2136-2008
- Company in FDA record
- AGFA Corp.
- FDA classification date
- Sep 18, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The "Film Calibration" setting on the printers was set to the default "OFF" position instead of "ON".
Source facts
- Action
- Consignees were notified by an Urgent Safety Notice letter on/about 05/16/2008. The letter provides the procedure the customer can follow to change the film calibration setting from "OFF" to "ON". The letter also provides information on how to get assistance from Agfa, if the customers prefer a service engineer perform the change. Consignees were instructed to send a copy of the letter or similar letter to their affected accounts. Agfa also offers to dispatch a service engineer to perform the parameter change.
- Initiated
- 2008-05-16
- Root cause
- Other
- Terminated
- 2008-10-10
- Distribution
- Worldwide Distribution --- USA including states of DE, FL, IL, KS, MA, MD, ME, MO, MT, NJ, NY, OH, OK, TX, UT, VA, WA, WV, and Puerto Rico and country of Canada.
- Recall number
- Z-2136-2008
- Recall status
- Terminated
- CDRH record ID
- 71475
- FDA classified
- 2008-09-18
- Recall event number
- 48462