Z-2136-2018: cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software co

Recall record Z-2136-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2136-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Jun 10, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.

Source facts

Action
On April 3, 2018, The Firm sent an Urgent Medical Device Correction letter via UPS Ground. Consignees were notified of the possible failure. Consignees were informed that a Roche representative would contact them to install a syringe plunger guide for the affected syringes to resolve the issue. In the meantime, consignees were instructed to only measure assays that do not require prewash steps and assays that are not impacted by missing prewash step until the syringe plunger guide is installed on the affected module. If the customer identifies broken syringe plungers on their device, the notice instructed them to contact the Roche Support Network Customer Support Center at 1-800-428-2336, available 24 hours a day, 7 days a week. Customers were asked to complete a customer response form and return it to the firm.
Initiated
2018-04-03
Root cause
Process design
Terminated
2019-11-21
Distribution
US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.
Recall number
Z-2136-2018
Recall status
Terminated
CDRH record ID
163480
FDA record created
2018-06-10
Recall event number
79805

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗