Skip to content

Public FDA record

Z-2136-2019: MED EL Cochlear Implant System, Mi1000 MED EL CONCERTO, sold under the following implant types: a. Mil

Recall record Z-2136-2019 for MED-EL Elektromedizinische Gereate, Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2136-2019
Company in FDA record
MED-EL Elektromedizinische Gereate, Gmbh
FDA record creation date
Aug 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.

Source facts

Action
Med-EL initially notified customers who had not yet implanted the affected devices on about 10/16/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to not implant the device if still available in inventory. Customers who had implanted the device were notified on about 11/27/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to immediately inform affected clinics of the recall and the long term risk. MED-EL does not recommend a device explantation, as the device may work as intended for years before the electronics may become affected by humidity ingress. Ultimately, accumulation of internal moisture may cause complete failure of the device. Should any uncomfortable or painful sensation be experienced, please stop immediately using the device and contact MED-EL or your local support center for immediate support. MED-EL recommends regular implant checks at least once a year to check the implant status. Additionally the customers were requested to respond to the FSN and confirm receipt.
Initiated
2018-10-16
Root cause
Employee error
Terminated
2020-05-22
Distribution
Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
Recall number
Z-2136-2019
Recall status
Terminated
CDRH record ID
169504
FDA record created
2019-08-02
Recall event number
81683

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗