Reason stated in the source

During product complaint investigation, it was determined that during stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.

Source facts

Action
On 06/07/2021, the firm initiated a verbal notification regarding an "URGENT MEDICAL DEVICE RECALL" which informed customers that Earlier this week, Medtronic identified the potential for a specific subset of 20Fr Endurant II/IIs Stent Graft Systems to have the potential for the Radiopaque Marker to detach during deployment. There is evidence that devices that used one batch of graft cover subassemblies has a higher potential to exhibit this failure mode. This is an acute failure of the delivery system, there are no concerns with previously implanted stent grafts. We are requesting that you immediately refrain from using this specific subset of products and place it in quarantine until further notice. You will be receiving an official recall notice with instructions for product return for credit or replacement. Your Sales Representative will assist you with this process. We apologize for the inconvenience; however, this action is necessary to ensure the utmost patient safety with the use of our products. On 06/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-Day delivery that restated the verbal notification. The Customer Notification Letter instructed customers to: 1)Return all unused affected devices via their sales representative or calling customer service at 1-888-283-7868 or email rs.aorticorders@medtronic.com. 2)Complete the enclosed Customer Confirmation Form and email to rs.cfqfca@medtronic.com. 3)Forward this notice to all those who need to be aware within their organization. For questions or assistance, contact Customer Service at 1-888-283-7868.
Initiated
2021-06-07
Root cause
Under Investigation by firm
Distribution
U.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.
Recall number
Z-2136-2021
Recall status
Open, Classified
CDRH record ID
188017
FDA record created
2021-07-23
Recall event number
88158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗