Open original FDA record ↗Recall record
- FDA identifier
- Z-2137-2008
- Company in FDA record
- Civco Medical Instruments Inc
- FDA classification date
- Sep 18, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A localization may be produced which has the incorrect moves, and therefore, could place the patient in the incorrect location.
Source facts
- Action
- On 06/16/08 Civco Medical Solutions issued an Urgent-Medical Device Correction letter to all its customers by certified mail. This letter states the circumstances in which the device works incorrectly, a temporary work around and the information that the new software version 6.7 will address the issue. The software will be released in August of 2008. Attached to the letter is an ISOLOC Device Correction Notification Customer Feedback form which is to be completed and returned acknowledging the receipt of the Device Correction Notification.
- Initiated
- 2008-06-18
- Root cause
- Software design
- Terminated
- 2011-03-31
- Distribution
- Worldwide Distribution --- USA including Puerto Rico and countries of Korea, Portugal, and United Kingdom.
- Recall number
- Z-2137-2008
- Recall status
- Terminated
- CDRH record ID
- 71476
- FDA classified
- 2008-09-18
- Recall event number
- 48463