Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
Source facts
Action
Med-EL initially notified customers who had not yet implanted the affected devices on about 10/16/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to not implant the device if still available in inventory.
Customers who had implanted the device were notified on about 11/27/2018 via telephone call and followed up with an "Urgent-Medical Device Field Safety Notice (FSN)." The notice instructed customers to immediately inform affected clinics of the recall and the long term risk. MED-EL does not recommend a device explantation, as the device may work as intended for years before the electronics may become affected by humidity ingress. Ultimately, accumulation of
internal moisture may cause complete failure of the device. Should any uncomfortable or painful sensation be experienced, please stop immediately using the device and contact MED-EL or your local support center for immediate support. MED-EL recommends regular implant checks at least once a year to check the implant status. Additionally the customers were requested to respond to the FSN and confirm receipt.
Initiated
2018-10-16
Root cause
Employee error
Terminated
2020-05-22
Distribution
Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
Recall number
Z-2138-2019
Recall status
Terminated
CDRH record ID
169506
FDA record created
2019-08-02
Recall event number
81683