Open original FDA record ↗Recall record
- FDA identifier
- Z-2138-2023
- Company in FDA record
- Beckman Coulter, Inc.
- FDA classification date
- Jul 12, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.
Source facts
- Action
- Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice on 05/31/2023 by email and mail. The notice explained the issue, its impact on the patient, and requested the following actions be taken: "Inspect the lot number set up on your instrument(s) and your current inventory with the corresponding container number (see Appendix A). Locate the container number on your diluent reagent 2D or Linear barcode label (see Appendix B). Contact Beckman Coulter Customer Technical Support Organization to replace the affected containers in your current inventory. Consult with your laboratory management to determine whether a retrospective review of results is necessary.
- Initiated
- 2023-05-31
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.
- Recall number
- Z-2138-2023
- Recall status
- Open, Classified
- CDRH record ID
- 200884
- FDA classified
- 2023-07-12
- Recall event number
- 92419