Skip to content

Public FDA record

Z-2138-2023: COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells

Recall record Z-2138-2023 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2138-2023
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Jul 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.

Source facts

Action
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice on 05/31/2023 by email and mail. The notice explained the issue, its impact on the patient, and requested the following actions be taken: "Inspect the lot number set up on your instrument(s) and your current inventory with the corresponding container number (see Appendix A). Locate the container number on your diluent reagent 2D or Linear barcode label (see Appendix B). Contact Beckman Coulter Customer Technical Support Organization to replace the affected containers in your current inventory. Consult with your laboratory management to determine whether a retrospective review of results is necessary.
Initiated
2023-05-31
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Bolivia (Plurinational State of), Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Indonesia, Kuwait, Mexico, Panama, Peru, Puerto Rico, Taiwan, Trinidad and Tobago, Uruguay, and Viet Nam.
Recall number
Z-2138-2023
Recall status
Open, Classified
CDRH record ID
200884
FDA classified
2023-07-12
Recall event number
92419

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗