Reason stated in the source
Table in LIS Specification depicting the association between Record ID and Numerical Result Label is incorrect. If a record ID is used to configure the system for mapping, results will come out nonsensical.
Source facts
- Action
- Recall initiated on April 11, 2008. A Product Correction letter and customer reply form were sent to all currently active CELL-DYN Sapphire Customers. The letter requests that users be certain that the transmission of test results receipt by the host interface is mapped according to the Numerical Result Label, when configuring their CELL-DYN Sapphire Hematology Analyzer for results transmission to a Clinical Laboratory Information System or Middleware. Affected product was distributed to a total of 4 distributors. However, notification was made to all 154 customers that received the CELL-DYNE Sapphire product. If the customer or the health care providers that they serve have any questions regarding this information, U.S. customers should call Customer Support at 1-877-4ABBOTT.
- Initiated
- 2008-04-11
- Root cause
- Software change control
- Terminated
- 2010-03-12
- Distribution
- Nationwide Distribution.
- Recall number
- Z-2139-2008
- Recall status
- Terminated
- CDRH record ID
- 70261
- FDA classified
- 2008-08-21
- Recall event number
- 47801