Skip to content

Public FDA record

Z-2139-2024: Aeos Robotic Digital Microscope, Product Code: PV010

Recall record Z-2139-2024 for Aesculap Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2139-2024
Company in FDA record
Aesculap Inc
FDA classification date
Jun 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Source facts

Action
On May 13, 2024, URGENT MEDICAL DEVICE FIELD CORRECTION letters were mailed to customers. Actions to be taken by the customer: 1. Review the Device Correction Notification in its entirety and ensure that all users in your organization of the device and other concerned persons are informed about this voluntary product correction. If you are a distributor, please forward this correction notification to your customers. The correction is to be extended to the end user/customer level. 2. Determine your current inventory of the affected material within your facility, cease use and quarantine product subject to correction. Do not destroy any affected product. (Once your completed Urgent Medical Device Form is received, a representative will be in contact to arrange for inspection and testing). 3. Utilizing the attached "Urgent Medical Device Correction Acknowledgement Form," record the total number of individual units in your posession. If you have no inventory, please enter zero (0) on the form. 4. Return the completed "Urgent Medical Device Correction Acknowledgement Form" to the Aesculap Inc. Quality Assurance Department by faxing the form to (610)849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). 5. Once we receive your acknowledgement Form, an Aesculap representative will contact you to arrange inspection and testing of your devices by our authorized specialists. Should you identify a problem or safety concern with the product, please report promptly by contacting our Post-market Surveillance Department at 1-833-425-1464. Additionally, adverse reactions of quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting Program wither online, by regular mail or fax.
Initiated
2024-05-13
Root cause
Process control
Distribution
US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.
Recall number
Z-2139-2024
Recall status
Open, Classified
CDRH record ID
207801
FDA classified
2024-06-12
Recall event number
94623

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗