Z-2140-2024: Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with

Recall record Z-2140-2024 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2140-2024
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Jun 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.

Source facts

Action
Baxter Healthcare began notifying consignees on about 05/29/2024 via " URGENT MEDICAL DEVICE RECALL" letter sent U.S.P.S first class mail. Patients, healthcare providers, wholesalers and distributors were instructed to ensure patients always have an alternate means of ventilator or oxygen therapy available, inspect the battery charger dongle for damage and replace immediately if damaged, and patients may continue to use the ventilator system once they inspect and confirm no damage to the dongle and battery is charging appropriately. Patients were also asked to observe any alarms of the ventilator system and notify Baxter Home Care Customer Service for any assistance, that Baxter will replace their ventilator device upon their next scheduled in-home visit with a clinical trainer, and to acknowledge receipt of the letter. Healthcare providers, wholesalers and distributors were requested to observe any alarms of the ventilator system and notify Baxter Acute Care Customer Service Support, 800-426-4224, option 2, between the hours of 8:00am and 5:00pm Central Time, Monday through Friday, for any assistance, arrange for the return of any impacted product and to acknowledge receipt of the letter. Additionally, they were asked to notify customers if further distributed. For general questions regarding this Medical Device Recall, Baxter Acute Care Customer Service Support at 800-426-4224, option 2, or Baxter Clinical Support at 800 397 9071.
Initiated
2024-05-29
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2140-2024
Recall status
Open, Classified
CDRH record ID
208045
FDA classified
2024-06-26
Recall event number
94701

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗