Open original FDA record ↗Recall record
- FDA identifier
- Z-2141-2008
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Aug 13, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Tip detachment: Boston Scientific has received complaints involving a detachment of the tip from the NexStent Carotid Stent delivery system. Potential health hazards resulting from this type of failure include increased procedure time, vessel wall injury, stroke and/or emergency surgery to remove the detached tip. Since the potential for the tip to detach occurs during delivery of the sten
Source facts
- Action
- Boston Scientific sent an urgent Voluntary Medical Device Recall letter, dated 06/06/08, to affected customers. The letter described the issue, identified affected product, requested to cease using or distributing the product and provided direction on how to return the recalled devices to Boston Scientific, Quincy, MA. A Reply Verification Tracking Form was asked to be returned to Boston Scientific. Since the issue is with the delivery system, patients already implanted with the stent are not affected.
- Initiated
- 2008-06-06
- Root cause
- Component design/selection
- Terminated
- 2011-02-16
- Distribution
- Worldwide Distribution: USA, Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, Italy, Kuwait, Luxembourg, Netherlands, Norway, Panama, Poland, Portugal, Romania, Spain, Switzerland, and Ukraine.
- Recall number
- Z-2141-2008
- Recall status
- Terminated
- CDRH record ID
- 71925
- FDA classified
- 2008-08-13
- Recall event number
- 48697