Z-2141-2024: US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic

Recall record Z-2141-2024 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2141-2024
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Jun 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

Source facts

Action
Fresenius Medical issued Urgent Medical Device Correction letter on 5-16-2024. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Water Systems Service Specialist will be contacting you to schedule an evaluation of your Water Treatment system and provide feedback on your system. For questions or concerns, please contact Fresenius Medical Care Technical Services at 800-227-2572, OPTION 4. " Please provide this letter to your Biomedical Technician " Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatchathttps://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. " Provide this Customer Notice to all those who need to be aware within your organization. " Complete and return the attached reply form.
Initiated
2024-05-16
Root cause
Process control
Distribution
US Nationwide distribution in the states of AL, IN, AK, WI.
Recall number
Z-2141-2024
Recall status
Open, Classified
CDRH record ID
207965
FDA classified
2024-06-13
Recall event number
94676

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗