Open original FDA record ↗Recall record
- FDA identifier
- Z-2142-2024
- Company in FDA record
- Ortho-Clinical Diagnostics, Inc.
- FDA classification date
- Jun 13, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for positively biased results due to signal reduction over shelf life.
Source facts
- Action
- Consignees were mailed an URGENT PRODUCT CORRECTION NOTIFICATION, dated 5/14/24. In this notice consignees are instructed to discontinue using, render unusable, and discard affected product. The provided notification is to be retained and posted by each VITROS System until the issue has been resolved; copies of the recall notification are to be forwarded if affected product was distributed outside your facility. Consignees are to complete and return the provided Confirmation of Receipt form no later than 5/22/24. Consignees with any questions are to contact Quidel Ortho's Global Services Organization at 1-800-421-3311.
- Initiated
- 2024-03-01
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MT, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WI, WV and the countries of Bermuda, Colombia, Singapore.
- Recall number
- Z-2142-2024
- Recall status
- Open, Classified
- CDRH record ID
- 207978
- FDA classified
- 2024-06-13
- Recall event number
- 94679