Z-2143-2016: enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and

Recall record Z-2143-2016 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2143-2016
Company in FDA record
Ortho-Clinical Diagnostics
FDA classification date
Jul 8, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The IFU states the interpretation term for these numerical result values is Retest?. The difference in terminology exists for US Markets only.

Source facts

Action
Ortho Clinical Diagnostics sent an "IMPORTANT PRODUCT CORRECTION NOTIFICATION" Customer Letter dated May 16, 2016, to all affected customers in the US with enGen Systems via FedEx or US Priority Mail (for Post Office Box addresses) or by e-mail to subscribers of ORTHO PLUS e-Communications to advise them of the issue and the required actions . Required Actions: 1. Use the numerical test results to assess actions to be taken for these assays, per the IFU. 2. Post this notification by your enGen System or with your user documentation. 3. Complete and return the Confirmation of Receipt form by 25 May 2016. with the Instructions for Use (IFU) directions). If you have any questions regarding this notification, please call the Ortho Technical Solutions Center at any time at 1-800-421-3311.
Initiated
2016-05-16
Root cause
Software design
Terminated
2018-06-08
Distribution
Nationwide
Recall number
Z-2143-2016
Recall status
Terminated
CDRH record ID
147261
FDA classified
2016-07-08
Recall event number
74354

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗