Z-2144-2023: Beckman Coulter DxA Automation System, DxA 5000, REF: B87352, IVD, 200240V, 16A, 50/60HZ, RxOnly

Recall record Z-2144-2023 for Beckman Coulter Biomedical GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2144-2023
Company in FDA record
Beckman Coulter Biomedical GmbH
FDA classification date
Jul 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect in the DxA Automation System may cause tests to be run on patient samples in tube types not compatible with the requested test, leading to erroneous patient results

Source facts

Action
On 06/12/2023, Beckman Coulter issued an Urgent Medical Device Recall Notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. Inspect tubes routed to error regions and confirm that the tube type was appropriate per the assay IFU for the test performed. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. 3. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: " Electronically, if you received this communication via email./" Manually, complete and return the enclosed Response Form. 4. If you have any questions regarding this notice, please contact our Customer Support Center or your local Beckman Coulter Representative: " From our website: http://www.beckmancoulter.com/" By phone: call 1-800-526-3821 in the United States. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2023-06-02
Root cause
Software design
Distribution
US: AL CA FL IL IN KY MI MN MO NE NJ NV NY OH PA SC TX WI WV OUS: Australia, Belgium, Canada, China, Croatia, Czechia, Egypt, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Kazakhstan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Saudi Arabia, Spain, Switzerland, United Kingdom,
Recall number
Z-2144-2023
Recall status
Open, Classified
CDRH record ID
201069
FDA classified
2023-07-14
Recall event number
92484

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗