Open original FDA record ↗Recall record
- FDA identifier
- Z-2146-2008
- Company in FDA record
- GE Healthcare
- FDA classification date
- Sep 26, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th
Source facts
- Action
- A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.
- Initiated
- 2008-07-01
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2011-12-11
- Distribution
- AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.
- Recall number
- Z-2146-2008
- Recall status
- Terminated
- CDRH record ID
- 73696
- FDA classified
- 2008-09-26
- Recall event number
- 49509