Z-2146-2016: QuickSlide MGS-80, Automated Mutli-Slide Gram Stainer/GramPRO-80. Systematically stains, rinses, decolor

Recall record Z-2146-2016 for Hardy Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2146-2016
Company in FDA record
Hardy Diagnostics
FDA classification date
Jul 8, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Discrepant culture and gram stain results, which could potentially lead to misdiagnosis or improper treatment of a patient.

Source facts

Action
The firm began notifying consignees via phone on 05/23/16. The firm also sent out Urgent Notification letters, dated June 15, 2016. The firm's phone script instructs customers to decrease the number of rinses on the slide and increasing the number of rinses in the chamber for better outcomes and performance with the instrument. Reducing the rinses on the slide may prevent bacteria from coming off the slide, and adding rinses to the empty chamber will help ensure that the chamber is clean before the next slide is introduced. The firm's customer notification letter recommended an immediate update to the automated rinse cycle to decrease the number of slide rinses and increase the number of empty chamber rinses to improve instrument performance. A step-by-step procedure for updating the rinse cycle settings was discussed via phone and should now be implemented. In addition to the changes, the firm is updating the GramPro 80 User Manual to reflect new recommendations, and it is strongly recommended that all users perform methanol fixation of slides for automated staining and discontinue the use of heat fixation to prevent cross contamination of the unit. Once updates to the User Manual have been completed, the firm will provide an updated revision for customer records. For further information or technical assistance, contact QuickSlide Technical Team at gibbsd@HardyDiagnostics.com.
Initiated
2016-05-24
Root cause
Under Investigation by firm
Terminated
2017-03-03
Distribution
U.S. Distribution to the following states: CA, TX, NE, OK, LA, NC, PA, AZ, NY, PA, VA, NM, MI, and WV.
Recall number
Z-2146-2016
Recall status
Terminated
CDRH record ID
147263
FDA classified
2016-07-08
Recall event number
74356

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗