Skip to content

Public FDA record

Z-2146-2019: Geistlich Bio-Oss Pen, Sterile cancellous granules in an applicator, 0.25-1.mm 0.5g, For single use only,

Recall record Z-2146-2019 for Geistlich Pharma North America, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2146-2019
Company in FDA record
Geistlich Pharma North America, Inc.
FDA classification date
Aug 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.

Source facts

Action
On April 12, 2019, Geistlich Pharma AG issued URGENT MEDICAL DEVICE CORRECTION NOTICES and Acknowledgement and Receipt Forms to customers via courier service. Customers were advised to take the following actions: 1) When using the device, please assure that the green cap is removed without force, and when removing the filter cap, the cap is turned to the left to loosen as overtightening could damage the cap. 2) Should you observe a similar plastic fragment, DO NOT use the product. Please retain the product and notify the firm by calling 855-799-5500 to arrange a return and replacement. 3) Customers requesting return of product call 855-799-5500 4) Complete and return the Acknowledgement form via email to orders@geistlich-na.com 5) Notify all clinicians within your practice by sharing a copy of the notice.
Initiated
2019-04-12
Root cause
Under Investigation by firm
Terminated
2020-10-21
Distribution
US Nationwide distribution.
Recall number
Z-2146-2019
Recall status
Terminated
CDRH record ID
172148
FDA classified
2019-08-02
Recall event number
82636

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗