Public FDA record

Z-2147-2008: GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, mod

Recall record Z-2147-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2147-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 26, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th

Source facts

Action
A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.
Initiated
2008-07-01
Root cause
Radiation Control for Health and Safety Act
Terminated
2011-12-11
Distribution
AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.
Recall number
Z-2147-2008
Recall status
Terminated
CDRH record ID
73706
FDA classified
2008-09-26
Recall event number
49509

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗