Open original FDA record ↗Recall record
- FDA identifier
- Z-2147-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- May 6, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mistreatment risk-- when a user defined formula is used to calculate a value in a vascular exam, data is switched laterally when it is internally transferred and displayed on the summary report within the system.
Source facts
- Action
- A field correction was initiated and a Customer Safety Advisory Notification was sent out on 7/29/09. The letter identified the affected product and stated that there was a potential for user defined calculations to be incorrect within the vascular package when using the listed software versions of the product. The letter also discussed the steps that customers could take to avoid the potential risk of the issue. A mandatory update software was developed and will be installed free of charge. Customers should have been contacted to schedule a time for the installation. Questions were to be directed to the local service support person.
- Initiated
- 2009-07-24
- Root cause
- Software design
- Terminated
- 2011-05-09
- Distribution
- Worldwide Distribution -- Austria, Brazil, China, Columbia, Germany, Denmark, Egypt, Great Britain, Honduras, Ireland, India, Italy Japan, South Korea, Mexico, Malaysia, Peru, Poland, Portugal, Romania, Russia, Sweden, Singapore, Thailand, Turkey, Taiwan, United States, Uruguay, Venezuela, and South Africa.
- Recall number
- Z-2147-2011
- Recall status
- Terminated
- CDRH record ID
- 84543
- FDA classified
- 2011-05-06
- Recall event number
- 53051