Skip to content

Public FDA record

Z-2147-2014: Boston Scientific ENDOTAK RELIANCE SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO.

Recall record Z-2147-2014 for Boston Scientific CRM Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2147-2014
Company in FDA record
Boston Scientific CRM Corp
FDA classification date
Aug 5, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A review of manufacturing test records for Boston Scientific ENDOTAK RELIANCE implantable leads revealed a suspected test data recording error. Specifically, some test results were recorded as "failed" without any other indication of failure. There are no reported injuries from the devices.

Source facts

Action
Sales representatives hand delivered a Boston Scientific "Medical Device Retrieval" Letter dated June 19, 2014 or June 20, 2014 when they were retrieving the devices from the hospital shelves. The letter was addressed to Hospital Administrator. The letter described the problem, product being recalled and the retrieval of the device by Boston scientific sales representative. For further questions they can contact their local sales representative or Boston Scientific International Technical Services. Physicians were contacted via telephone starting on July 6, 2014. A Boston Scientific letter dated July 8, 2014 followed the telephone conversation. For further information physicians can contact their local Boston Scientific representative or Technical Services at 1-800-227-3422.
Initiated
2014-06-19
Root cause
Software Manufacturing/Software Deployment
Terminated
2014-12-03
Distribution
Worldwide Distribution -US (nationwide) in the states of OH, MN and country of France
Recall number
Z-2147-2014
Recall status
Terminated
CDRH record ID
128470
FDA classified
2014-08-05
Recall event number
68700

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗