Open original FDA record ↗Recall record
- FDA identifier
- Z-2147-2019
- Company in FDA record
- Geistlich Pharma North America, Inc.
- FDA classification date
- Aug 2, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.
Source facts
- Action
- On April 12, 2019, Geistlich Pharma AG issued URGENT MEDICAL DEVICE CORRECTION NOTICES and Acknowledgement and Receipt Forms to customers via courier service. Customers were advised to take the following actions: 1) When using the device, please assure that the green cap is removed without force, and when removing the filter cap, the cap is turned to the left to loosen as overtightening could damage the cap. 2) Should you observe a similar plastic fragment, DO NOT use the product. Please retain the product and notify the firm by calling 855-799-5500 to arrange a return and replacement. 3) Customers requesting return of product call 855-799-5500 4) Complete and return the Acknowledgement form via email to orders@geistlich-na.com 5) Notify all clinicians within your practice by sharing a copy of the notice.
- Initiated
- 2019-04-12
- Root cause
- Under Investigation by firm
- Terminated
- 2020-10-21
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2147-2019
- Recall status
- Terminated
- CDRH record ID
- 172149
- FDA classified
- 2019-08-02
- Recall event number
- 82636