Open original FDA record ↗Recall record
- FDA identifier
- Z-2150-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- May 6, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The weld connecting the ABVS arm to the system base has the potential to fail, causing the arm to fall off of the unit.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. initiated the recall via phone calls and e-mail on March 28, 2011, to all affected customers, informing them of the affected product and providing instructions on the recall. Three units S/N 250171, 250173, and 250177 were to be removed and shipped back to Siemens. Siemens Service personnel were to be dispatched to the customer sites to perform validated stress test on units S/N 250157, 250162, 250163, 250164, 250166, 250167, and 250169 at customer sites. For questions regarding this recall call 650-694-5993.
- Initiated
- 2011-03-28
- Root cause
- Process change control
- Terminated
- 2011-05-06
- Distribution
- Worldwide Distribution - USA, Portugal, Germany, Russia, Estonia, Brazil, France, and Saudi Arabia.
- Recall number
- Z-2150-2011
- Recall status
- Terminated
- CDRH record ID
- 99480
- FDA classified
- 2011-05-06
- Recall event number
- 58504