Skip to content

Public FDA record

Z-2150-2024: TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-fo

Recall record Z-2150-2024 for Outset Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2150-2024
Company in FDA record
Outset Medical, Inc.
FDA classification date
Jun 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or the associated connector power entry module, which may lead to transient delay in treatment or blistering to the user.

Source facts

Action
On 4/30/24, correction notices were sent to customers asking them to do the following: The firm has enclosed the replacement power cord (PN-0008460) with this communication based on the number of affected devices at your facility. Disseminate this information to appropriate personnel within your facility for processing and distribution. This notice should be shared with any organization where the potentially affected devices have been transferred. Follow instructions to replace the power cords. Dispose of the affected power cords. Complete and return the acknowledgement and receipt form via email to FieldActionAdvisory@OutsetMedical.com If you have questions or notice any damage around the power cord, please contact us at support@outsetmedical.com or 844-MYTABLO (844-698-2256) and request an inspection from a Field Service Engineer.
Initiated
2024-04-30
Root cause
Device Design
Distribution
US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK, MA, WV, IN, KY, AZ, LA, NE, KS, NJ, CT, NM, CO, SC, TX, PA, NY, OH.
Recall number
Z-2150-2024
Recall status
Open, Classified
CDRH record ID
207935
FDA classified
2024-06-14
Recall event number
94654

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗