Reason stated in the source
Calibration and Quality Control outside of established range resulting from an increase in initial OD readings with the reagent contained within the kits.
Source facts
- Action
- Genzyme notified accounts by Fed'X on 4/14/08 via an Imprtant Field Correction letter, requesting users to discard kits and follow the "User Update" instructions previously provided to minimize calibration error. Customers are to complere and sign the Confirmation of Notification form and fax it back to the firm regardless if they have any affected product or not. Also, requires that the customer confirms that the REagent OD limits for First and Last H are set at 0.15 for Lipase on the Olympus AU400/400e, AU600, AU640/640e and AU2700/AU5400 Systems.
- Initiated
- 2008-04-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-09-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-2151-2008
- Recall status
- Terminated