Open original FDA record ↗Recall record
- FDA identifier
- Z-2151-2012
- Company in FDA record
- Linvatec Corp. dba ConMed Linvatec
- FDA classification date
- Aug 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm when it should have been 3.5mm.
Source facts
- Action
- The recall is being conducted to the end user level. ConMed Linvatec Urgent Medical Device Recall Notifications were sent by FedEx next day delivery or equivalent method to notify all direct consignees. The notifications identified the affected product along with the reason for the recall and health risk. All accounts were asked to immediately check their facility for the affected product, stop using any of the product with specified lot number, and to segregate the product. The letter also asked customers to return all affected product to ConMed Linvatec and respond using a Reply Form, even if there is no affected product available. If customers have any additional questions, they should contact the Customer Service department at 800-237-0169.
- Initiated
- 2012-07-06
- Root cause
- Process change control
- Terminated
- 2013-11-21
- Distribution
- Worldwide Distribution -- US and Korea.
- Recall number
- Z-2151-2012
- Recall status
- Terminated
- CDRH record ID
- 111312
- FDA classified
- 2012-08-06
- Recall event number
- 62665