Skip to content

Public FDA record

Z-2151-2017: Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold i

Recall record Z-2151-2017 for Abbott Vascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2151-2017
Company in FDA record
Abbott Vascular
FDA classification date
Jun 1, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.

Source facts

Action
Abbott Vascular sent an Urgent Medical Device Recall and Hazard Alert letter dated April 26, 2017, to affected customers to inform them that the action is in response to recent concerns over data from some studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent. Customers were informed of the actions to be taken. Customers were instructed to stop using these devices immediately, advise BVS patients to follow the recommendations for DAPT prescribed by their health care provider. Advise patients experiencing new cardiac symptoms to seek clinical care, review their inventory and complete the attached Facsimile Reply Form and display and share this notification with other relevant personnel within their organization. Customers with questions are instructed to contact their local Abbott Vascular Representative or Customer Service department.
Initiated
2017-04-26
Root cause
Other
Terminated
2017-11-16
Distribution
Australia
Recall number
Z-2151-2017
Recall status
Terminated
CDRH record ID
155009
FDA classified
2017-06-01
Recall event number
77131

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.