Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.
Source facts
Action
Fresenius issued on 05/29/19 The IMPORTANT MEDICAL DEVICE INFORMATION notification explains to the Home Therapy Nurses/Clinics to review the new flyer with each of their APD patients who use a Liberty Cycler.
For the patient notification, it will encourage them to use and follow the attached instructional flyer during set up for their daily dialysis treatment. The notification will have a copy of the new step-by-step instructional flyer. Included is a Reply Form for confirmation the consignee read and understands the notice sent to them.
Initiated
2019-05-29
Root cause
Use error
Terminated
2020-11-19
Distribution
US Nationwide Distribution
Recall number
Z-2151-2019
Recall status
Terminated
CDRH record ID
173897
FDA record created
2019-08-05
Recall event number
83164