Z-2151-2025: Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSP

Recall record Z-2151-2025 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2151-2025
Company in FDA record
Covidien
FDA classification date
Jul 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Source facts

Action
On 6/23/2025 correction notices were distributed to customers who where informed: Per Operator s Manual (http://manuals.medtronic.com): 1) Device alarm volume needs to be set to a level appropriate to be heard throughout the care environment at all times of day or night. Do not silence or decrease the volume of the audible alarm if patient safety could be compromised. 2) If the sensor comes off the patient, or no perfusion is detected, the device will notify 'sensor off' (low priority alarm) every 16 seconds and display a steady yellow visual indicator. If a low priority alarm immediately follows a higher priority alarm, a user may not recognize the higher priority clinical condition. 3) Monitors should be set to Standard Mode when in use in a hospital or hospital-type environment by trained medical personnel and set to Homecare Mode when a lay person will be using the monitor outside of a hospital or other professional care setting. With the monitor not in Homecare Mode, alarms could be silenced/muted and alarm limits altered In Clinical Facility Environments be sure the PM100N model is set to Standard Mode for general use and only set to Sleep Study Mode when conducting a sleep study. Device should be set to Homecare Mode prior to use. Firm's Technical Service at 800-255-6774 (option 1, then 2) can provide a four-digit password required to set a monitor to Homecare Mode. Provide the Home Use Guide to home caregivers. Clinicians should communicate the recall information to home use patients and caregivers. Pass on/post this notice for all those who need to be aware within your organization and to any organization where the product has been transferred/distributed. Complete and return the Customer Confirmation Form to rs.gmbmitgfca@medtronic.com Firm will provide additional information when finalized. Questions - Contact firm's Customer Service at 800-962-9888, Option 2.
Initiated
2025-06-23
Root cause
Labeling design
Distribution
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.
Recall number
Z-2151-2025
Recall status
Open, Classified
CDRH record ID
213817
FDA classified
2025-07-21
Recall event number
96862

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗