Public FDA record

Z-2152-2023: Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-0

Recall record Z-2152-2023 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2152-2023
Company in FDA record
Datascope Corp.
FDA classification date
Jul 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.

Source facts

Action
On June 5, 2023, the firm notified customers of the recall via "URGENT MEDICAL DEVICE CORRECTION" letters. Letters covered two issues: Issue 1 - Failure to charge battery (Covered by RES 92287) Issue 2 - Unexpected shutdown (Covered by this RES event) User Actions: Issue 2: Unexpected Shutdown due to failure of Tantalum Capacitors. Should loss of power occur, therapy will be interrupted and the console will need to be exchanged with another (if available) to resume therapy. If your device remains inoperable, please contact your service representative. It is recommended for users to have an alternate IABP on hand in the event that this issue may occur. The firm is developing a hardware correction to address this issue. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. ET
Initiated
2023-06-05
Root cause
Component design/selection
Distribution
Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GHANA, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM, YEMEN, ZAMBIA
Recall number
Z-2152-2023
Recall status
Open, Classified
CDRH record ID
200331
FDA classified
2023-07-20
Recall event number
92335

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗